Advantech PCI-1720U Bedienungsanleitung Seite 37

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 38
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 36
31 Appendix A
A.2 Unipolar Output Calibration
1. Select an appropriate output range for the channel to be calibrated.
2. Set all digital input codes to 0. Then adjust VRn (n = 5, 6, 7, 8
depending on what channel is to be calibrated, see prior page) until
your voltmeters reading is 0.000 volts.
3. Set all digital input codes to 1 (see Unipolar Binary Code Table in
Section 4.3). Now, adjust VRn (n = 1, 2, 3, 4 depending on what
channel is to be calibrated, see prior page) until your voltmeter
shows a reading equal to the output voltage shown in the Unipolar
Binary Code Table in Section 4.3.
4. Repeat steps 2 and 3 until both of them are satisfied.
A.3 Bipolar Output Calibration
1. Select an appropriate output range for the channel to be calibrated.
2. Set all digital input codes to 100000000000. Adjust VRn (n = 5, 6,
7, 8 depending on what channel is to be calibrated, see prior page)
until your voltmeters reading is 0.000 volts.
3. Set all digital input codes to 1 (see Bipolar Binary Code Table in
Section 4.3). Now, adjust VRn (n = 1, 2, 3, 4 depending on what
channel is to be calibrated, see prior page) until your voltmeter
shows a reading equal to the output voltage shown in the Bipolar
Binary Code Table in Section 4.3.
4. Repeat steps 2 and 3 until both of them are satisfied.
Seitenansicht 36
1 2 ... 32 33 34 35 36 37 38

Kommentare zu diesen Handbüchern

Keine Kommentare